C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® PEDIATRIC ALL-SILICONE FOLEY CATHETER Back to Search Results
Model Number 165810
Medical Device Problem Codes Inadequacy of Device Shape and/or Size (1583); Material Too Soft/Flexible (4007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the foley catheters were leaking and dislodging every time they inserted into the patient.This was a transgender patient converting from a female to a male.Stated that they need a stiffer or firmer foley for this patient because they need to keep the urethra open for an extended period of time because there were multiple reconstruction surgeries with this transition.Per photo sample, it was found that the foley catheter balloons were asymmetrical.The affected product catalog numbers were 165810 and 165812.
 
Additional Manufacturer Narrative
As per review, this event is not reportable.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the foley catheters were leaking and dislodging every time they inserted into the patient.This was a transgender patient converting from a female to a male.Stated that they need a stiffer or firmer foley for this patient because they need to keep the urethra open for an extended period of time because there were multiple reconstruction surgeries with this transition.Per photo sample, it was found that the foley catheter balloons were lopsided which was leading to the issues that customer was expressing.The affected product catalog numbers were 165810 ( batch# ngft5821) and 165812.(batch# ngez0706 and # ngfw3722).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® PEDIATRIC ALL-SILICONE FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key14029570
Report Number1018233-2022-02065
Device Sequence Number4531783
Product Code KOD
UDI-Device Identifier00801741029721
UDI-Public(01)00801741029721
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K821745
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/01/2026
Device Model Number165810
Device Catalogue Number165810
Device Lot NumberNGFT5821
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/11/2022
Supplement Date Received by Manufacturer09/17/2022
Initial Report FDA Received Date04/06/2022
Supplement Report FDA Received Date09/22/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-