STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number TMICL13.2
Medical Device Problem Codes Inadequacy of Device Shape and/or Size (1583); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 03/16/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
The reporter indicated the surgeon implanted a ticl toric implantable collamer lens.When inserting the lens, the surgeon thought the lens may have been inserting upside down, but when the lens unfolded in the eye, the lens was properly under the iris in the correct orientation.At one day post-op, the eye had a very shallow anterior chamber, inappreciable vault and unrecordable intraocular pressure (iop).The surgeon realized there was a small wound leak and he repaired it with one suture.The anterior chamber deepened, there was appreciable vault and iop was low but recordable.The lens remained implanted.Additional information has been requested but none has been forthcoming.
 
Additional Manufacturer Narrative
Additional data: b5: the surgeon implanted a 13.2mm tmicl13.2 implantable collamer lens, -13.50/+1.0/089 (sphere/cylinder/axis) in the patients left eye (os) on (b)(6) 2022.The lens had a low vault.The cause of the event was unknown.The patient is doing great.H6 - work order search: no similar complaint was reported for units within the same lot.Corrected data: b4 and g3: 16-mar-2022.Claim # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key14030182
Report Number2023826-2022-01105
Device Sequence Number14627354
Product Code QCB
UDI-Device Identifier00841542109505
UDI-Public00841542109505
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2023
Device Model NumberTMICL13.2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/17/2022
Supplement Date Received by Manufacturer05/04/2022
Initial Report FDA Received Date04/06/2022
Supplement Report FDA Received Date05/27/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age31 YR
Patient SexMale
-
-