Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
H3: as found condition (condition of returned device): the solitaire fr stent was returned for analysis within a shipping box and within an opened solitaire fr stent outer carton.Visual inspection/damage location details: as per the solitaire fr stent ifu (instructions for use) the 6-40 stent must be used with a micro catheter with a minimum id (inner diameter) of 0.027in.Per the phenom-27 micro catheter product labeling, the id is 0.027¿.Therefore, the solitaire fr stent was found to be compatible for use with the phenom-27 micro catheter.The solitaire fr stent device was returned within phenom-27 catheter.The solitaire pushwire was found to be extending ~9.5cm from hub and ~13.9cm from distal tip.The solitaire fr stent device was pushed out from within the phenom catheter distal tip without issue.(pli-10) no bends or kinks were found with the solitaire fr pushwire or marker coil; however, the pushwire was found to be broken at ~182.4cm from proximal end.No corrosion was found on the broken end of the pushwire.Visual inspection of the detachment zone revealed no electrical etching.The detachment zone was found to be damaged.The stent was found to be still attached to the pushwire.The stent non-working (tear drop) was found to be damaged (kinked).The working length struts were found to be in good condition.The finger markers were found to be in good condition.(pli-20).No flash or voids molded were observed with the hub.The catheter body was found to be accord ioned at ~22.0cm or ~0.7, accordioned at ~16.0cm for ~7.8cm, accordioned at ~5.3cm for ~1.1cm from distal tip.No damages or irregularities were found with the marker band.The distal tip was found to be flattened.The catheter total length was measured and found to be ~165.9cm and the usable length was found to be ~159.5cm.No other anomalies were observed.Testing/analysis (including sem reports): the solitaire fr stent was sent out for sem (scanning electron microscopy) analysis for failure analysis of the broken pushwire.Per the sem analysis report, fatigue cracking initiated from the wire surface is visible.The rest of the wire failed via overload.No other anomalies were observed.Conclusion: based on the device analysis and reported information, the customer¿s report of ¿stent stuck in catheter hub¿ was unable to be confirmed.Possible causes for ¿stent stuck in catheter hub¿ are, damaged catheter, or sheath not fully seated in hub.It is likely the cause of stent stuck in catheter hub is incompatible catheter.The investigation determined that this event was similar to an event that had already been investigated, and another investigation is not necessary.Based on the analysis findings, the customer¿s report of ¿catheter resistance at hub¿ were unable to be confirmed and the root cause could not be determined.It is possible that introducer sheath did not sit securely inside hub, or user pulls back on/torques wire during insertion may have contributed to the reported issue.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|