| Model Number |
S722 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
Unspecified Mental, Emotional or Behavioural Problem (4430)
|
| Date of Event |
01/09/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The product has been received for analysis.This report will be updated upon completion of analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient implanted with this device presented to the emergency room feeling uneasy.Upon interrogation it was identified the device was operating in safety mode.New information received indicates that the device was explanted and will be returned for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.A high current drain was detected, but the battery still supported full device function.Detailed analysis confirmed the identified high current drain was a result of compromised capacitors.This resulted in the observed premature battery depletion.It was determined that the capacitors were compromised due to the presence of excess hydrogen in the device case.Boston scientific issued a field safety notice regarding a subset of devices in the accolade pacemaker family that have a potential of exhibiting this behavior.This particular device is included in the hydrogen induced premature depletion advisory population.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient implanted with this device presented to the emergency room feeling uneasy.Upon interrogation it was identified the device was operating in safety mode.New information received indicates that the device was explanted and will be returned for evaluation.
|
| |
|