| Model Number |
D021 |
| Medical Device Problem Codes |
Failure to Capture (1081); High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/31/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The local area field representative was contacted for additional information.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) and right ventricular (rv) defibrillation lead had triggered a red alert due to high out-of-range pace impedance measurements greater than 3,000 ohms.In addition, there was rv loss of capture (loc).This icd and rv lead remain in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a malfunction or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.Investigation has determined that this type of event is likely the result of an intermittent high impedance condition associated with the device spring contact and lead terminal ring.A design enhancement was implemented in 2020 to stabilize the electrical connection between the spring contact and the lead terminal.-- the local area field representative was contacted for additional information.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) and right ventricular (rv) defibrillation lead had triggered a red alert due to high out-of-range pace impedance measurements greater than 3,000 ohms.In addition, there was rv loss of capture (loc).This icd and rv lead remain in service.No adverse patient effects were reported.
|
| |
|