• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONFORMIS, INC. ITOTAL IDENTITY CR; TOTAL KNEE REPAIR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONFORMIS, INC. ITOTAL IDENTITY CR; TOTAL KNEE REPAIR SYSTEM Back to Search Results
Model Number TCR101111201
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 03/07/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that replacement poly inserts have been requested for revision surgery.Reason for revision is an infection of unknown cause.Primary surgery occurred on (b)(6) 2021.Revision surgery is planned for (b)(6) 2022.
 
Additional Manufacturer Narrative
It was reported that replacement poly inserts have been requested for revision surgery.Reason for revision is an infection of unknown cause.Primary surgery occurred on (b)(6) 2021.Revision surgery is planned for (b)(6) 2022.The device dhr was reviewed and the device was manufactured to specification.All sterilisation requirements were met.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ITOTAL IDENTITY CR
Common Device Name
TOTAL KNEE REPAIR SYSTEM
Manufacturer (Section D)
CONFORMIS, INC.
600 technology park drive
billerica MA 01821
Manufacturer (Section G)
CONFORMIS, INC.
600 research drive
wilmington MA 01887
Manufacturer Contact
kara johnson
600 technology park drive
billerica, MA 01821
MDR Report Key14034693
Report Number3004153240-2022-00052
Device Sequence Number6298675
Product Code JWH
UDI-Device IdentifierM572TCR1011112011
UDI-PublicM572TCR1011112011
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K210809
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberTCR101111201
Device Catalogue NumberTCR-101-1112
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2022
Initial Report FDA Received Date04/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age60 YR
Patient SexUnknown
-
-