| Catalog Number |
2R8402 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device manufacturer address 1: (b)(4).The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported the tubing of a clearlink system non-dehp solution set leaked.The event occurred during an unknown process step.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.The device underwent functional testing including, priming, pressure and clear passage testing; and a leak was observed between the assembly of the tubing and into the drip chamber.Dimensional testing was according to product specification.The reported condition was verified.The cause of the condition was due to improper manual assembly.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|