It was reported that, during an acl reconstruction procedure, the biosure screw broke inside the patient.All broken pieces were removed from the patient user grasper and suction.The procedure was completed with a non-significant delay using a smith and nephew back up device used in the originally drilled bone hole.No further complications were reported.
|
|
H10: internal complaint reference: case (b)(4).H6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.It was determined the device contributed to the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the polymer found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.A review of the customer provided image found a screw that has fractured into multiple pieces.There is debris on the device.Without responses to the documentation/information requests, the reported event could not be further assessed.According to the report, all the broken pieces were removed from inside of the patient using grasper and suction.It was reported, the procedure was completed with a non-significant delay using a smith and nephew back up device used in the originally drilled bone hole.Since no harm has been alleged to this patient and no other complications were reported, no further clinical/medical assessment is warranted at this time.Should any additional relevant medical information be provided, this case would be re-assessed.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|