H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one sidekick crossing support catheter loaded onto the crosser catheter was returned for evaluation.Sidekick was noted to be warped and bunching at the distal tip.An attempt was made to remove the sidekick loaded onto the crosser but was unsuccessful.Based on the findings, the investigation is confirmed for the reported retraction problem, device-device incompatibility and material deformation issues.A definitive root cause for the reported retraction problem, device-device incompatibility and material deformation issues could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 09/2023), h11: h6 (result, conclusion), h11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|