• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRNY II CR ISRT XLPE LT SZ 3-4 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRNY II CR ISRT XLPE LT SZ 3-4 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 74025641
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 03/15/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after total knee replacement was performed on (b)(6) 2022, the patient experienced an aseptic loosening.This adverse event was treated with a revision surgery on (b)(6) 2022.Patient's current health status is unknown.
 
Additional Manufacturer Narrative
(b)(4).
 
Additional Manufacturer Narrative
The device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, per complaint details, a revision was performed due to aseptic loosening.Reportedly, the cr tka was replaced with a lgn revision.The requested clinical documentation has not been provided as of the date of this medical investigation and the current patient status remains unknown.Without the requested information, further root cause analysis of the reported event cannot be performed.The patient impact beyond the reported aseptic loosening and subsequent revision cannot be determined based on the limited information provided.No further medical assessment can be rendered at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for knee systems revealed that loosening has been identified as a probable adverse effect.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors and/or some potential probable causes that could contribute to the reported event have been identified as abnormal motion over time, bone degeneration, fit/sizing issue, lack of ingrowth, and/or traumatic injury.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II CR ISRT XLPE LT SZ 3-4 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14040885
Report Number1020279-2022-01613
Device Sequence Number17710669
Product Code JWH
UDI-Device Identifier00885556422533
UDI-Public00885556422533
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K121443
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number74025641
Device Catalogue Number74025641
Device Lot Number19EM25191
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/15/2022
Supplement Date Received by Manufacturer05/03/2022
Initial Report FDA Received Date04/07/2022
Supplement Report FDA Received Date05/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/29/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention; Hospitalization;
Patient SexUnknown
-
-