Model Number 55-0513
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Inadequate Osseointegration (2646)
Date of Event 03/22/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The patient was revised due to metal on metal with ultima implants.There was metalosis present and corrosion on neck taper.Surgeon was successful removing the srom stem rather easily and had to work a bit to take out sleeve.Replaced srom with arcos from zimmer.Replaced metal liner with pinnacle dual mobility metal liner.There was loosening at bone to implant interface.Doi: (b)(6) 2007.Dor: (b)(6) 2022.Affected side: right hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide lot specific complaint database search, or manufacturing record evaluation, was not possible as the required lot number was not provided.The information received will be retained for trend analysis, post market surveillance, or other events within the quality system.Device history lot: a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
S-ROM HIP SYSTEM : HIP FEMORAL SLEEVE
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14041822
Report Number1818910-2022-06316
Device Sequence Number12334334
Product Code LPH
UDI-Device Identifier10603295175230
UDI-Public10603295175230
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P070026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number55-0513
Device Catalogue Number550513
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/22/2022
Supplement Date Received by Manufacturer04/21/2022
Initial Report FDA Received Date04/07/2022
Supplement Report FDA Received Date04/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PINNACLE MTL INS NEUT36IDX50OD; S-ROM*SLEEVE PRX ZTT, 16D-SM; S-ROM*SLEEVE PRX ZTT, 16D-SML; UNK HIP FEMORAL HEAD METAL SROM
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-