Model Number 48280233
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
A company representative reported that the tip of an es2 k-wire fractured during surgery and was left in the patient's bone.The procedure was completed successfully with no surgical delay and no adverse consequence to the patient.
 
Additional Manufacturer Narrative
Device was discarded by hospital.
 
Event or Problem Description
A company representative reported that the tip of an es2 k-wire fractured during surgery and was left in the patient's bone.The procedure was completed successfully with no surgical delay and no adverse consequence to the patient.
 
Additional Manufacturer Narrative
Visual, dimensional, material and functional analysis could not be performed as the device was not returned.Device history records could not be reviewed as a valid lot number was not provided and could not be obtained.Complaint history records were reviewed for this catalog number, and no adverse trends were observed.It was reported that the tip of an es2 k-wire fractured during surgery and the tip was left in the patient's bone.The procedure was completed successfully with no surgical delay and no adverse consequence to the patient.The es2 surgical technique was reviewed: "carefully determine the appropriate entry point and trajectory for the es2 screw.The entry point for percutaneous pedicle screws is approximately 4-5cm off mid-line with a more lateral trajectory." it is unknown if the k-wire was placed at a difficult angle or if a k-wire guide tube was used to insert the k-wire.Tip of the k-wire may have pinched while inserting awl/ screw, leading to increasing tension on the wire causing fracture."incorrect alignment of the k-wire and screw path may also cause pinching of the wire.".Limited information regarding the event was made available to stryker.Multiple attempts were made to obtain additional information, but no response was received.A root cause cannot be established with the information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ES2 LE NITINIOL KWIRE SHARP
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
Manufacturer (Section G)
STRYKER SPINE-FRANCE
zone industrielle de marticot
cestas 33610
FR   33610
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key14042820
Report Number0009617544-2022-00041
Device Sequence Number6298317
Product Code LXH
UDI-Device Identifier07613327266450
UDI-Public07613327266450
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number48280233
Device Catalogue Number48280233
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/15/2022
Supplement Date Received by Manufacturer06/10/2022
Initial Report FDA Received Date04/07/2022
Supplement Report FDA Received Date07/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-