MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAL; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-723LNAL
Medical Device Problem Codes Mechanical Problem (1384); Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/18/2019
Type of Reportable Event Malfunction
Event or Problem Description
Information received by medtronic indicated that the insulin pump had a motor error alarm.Customer also reported no delivery.Customer was able to clear the alarm and rewind the insulin pump.No harm requiring medical intervention was reported.The customer will discontinue to use the device.
 
Additional Manufacturer Narrative
Sw 2.5b.On the event date (b)(6) 2019.Customer returned pump for an alleged issue with no delivery alarm and motor error.Pump passed a21 error test, seat, rewind and basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and displacement test.No delivery alarm and no motor error alarm during testing.Pump history download using thds was successful.Unable to confirm reported event date & time and to verify whether alarm occurred during basal or bolus due to a47 alarms found on the history file which caused corrupted data.A47 alarms are due to the pump not having power for a long period of time.Unit was cut/open and inspected no moisture damage or component damage found inside the pump.The test p-cap/reservoir does lock into place.Motor passed motor test.The following were noted during visual inspection: cracked reservoir tube lip.Pump passed all required testing.No unexpected no delivery alarm and motor error alarm during testing.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAL
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*  00777-3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*   00777-3869
Manufacturer Contact
tricha miles
ceiba norte ind. park #50 road
juncos 00777--386
*   00777-3869
7635140379
MDR Report Key14043027
Report Number2032227-2022-176888
Device Sequence Number2239406
Product Code OYC
UDI-Device Identifier00643169478589
UDI-Public(01)00643169478589
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-723LNAL
Device Catalogue NumberMMT-723LNAL
Device Lot NumberA4723LNALJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2019
Initial Report FDA Received Date04/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/28/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age62 YR
Patient SexMale
Patient Weight168
-
-