| Catalog Number |
8065752082 |
| Medical Device Problem Codes |
Inability to Irrigate (1337); No Flow (2991)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
An ophthalmic surgeon reported irrigation failure and the solution did not come out during a cataract procedure.The fourth replacement product was obtained and the procedure was completed.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The customer did not retain the product lot information for this procedure pack, therefore the device history records traceable to the reported procedure pack could not be reviewed.Based on our current tracking, there are no adverse trends for this reported complaint.Irrigation failure was reported.Three used intrepid plus fluidics management system (fms) were visually inspected and no obvious defects were detected.The samples were tested using an infiniti console.The samples primed and tuned with the ultrasonic handpiece, the 0.9 millimeter (mm) air bypass system (abs) tip and infusion sleeve from lab stock successfully.Irrigation rate met specification.The root cause of the customer's complaint could not be established; the returned sample functioned per specification.This complaint has been reviewed and it is determined that no further actions will be pursed at this time as the sample met specifications.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|