|
As reported by the field, during a coil embolization of the anterior communicating artery (acom), the physician was not able to insert a galaxy g3 mini 1.5mm x 2.5cm coil (glm915025, s15171) or a galaxy g3 mini 1.5mm x 3cm coil (glm915030, l11005) into the hub of echelon14 (14/105-5092-150) tip straight microcatheter (medtronic).There was no patient injury.The procedure was delayed by 15 minutes or more due to the events.The procedure was successfully completed with another device.Additional information received indicated that there had been no excessive force applied to the devices.It was no necessary to remove concomitant devices with the complaint device.There were no device fragments that remained in the patient.The devices were used and prepped as per the instructions for use (ifu).Adequate flush was maintained through the devices.A target and axium coils (with bigger diameter) were used successfully with the same microcatheter (mc).The events did not result in a loss of cerebral target position.The mc was not damaged during manipulation of the coils or noticed to be damaged upon removal from the patient.Nothing was noted to be obstructing the mc.The introducer was flushed until liquid was visible at the distal end of the slit in the clear tube.The tip of the introducer was firmly installed into the hub of the microcatheter and locked with the rhv during coil advancement.Based on the product analysis of both devices received, the embolic coil of the galaxy g3 mini 1.5mm x 3cm was found damaged ((b)(4)) and the embolic coil of the galaxy g3 mini 1.5mm x 2.5cm ((b)(4)) was found stretched.
|
|
Product complaint # (b)(4).The purpose of this mdr submission is to report the findings of the device investigation.The embolic coil was found damaged, meeting regulatory reporting criteria.Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Procode: krd/hcg.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.This is one of two products involved with the complaint and the associated manufacturer report numbers are 3008114965-2022-00269 and 3008114965-2022-00270.Complaint conclusion: as reported by the field, during a coil embolization of the anterior communicating artery (acom), the physician was not able to insert a galaxy g3 mini 1.5mm x 2.5cm coil (glm915025, s15171) or a galaxy g3 mini 1.5mm x 3cm coil (glm915030, l11005) into the hub of echelon14 (14/105-5092-150) tip straight microcatheter (medtronic).There was no patient injury.The procedure was delayed by 15 minutes or more due to the events.The procedure was successfully completed with another device.Additional information received indicated that there had been no excessive force applied to the devices.It was no necessary to remove concomitant devices with the complaint device.There were no device fragments that remained in the patient.The devices were used and prepped as per the instructions for use (ifu).Adequate flush was maintained through the devices.A target and axium coils (with bigger diameter) were used successfully with the same microcatheter (mc).The events did not result in a loss of cerebral target position.The mc was not damaged during manipulation of the coils or noticed to be damaged upon removal from the patient.Nothing was noted to be obstructing the mc.The introducer was flushed until liquid was visible at the distal end of the slit in the clear tube.The tip of the introducer was firmly installed into the hub of the microcatheter and locked with the rhv during coil advancement.A non-sterile galaxy g3 mini 1.5mm x 3cm that the core wire was observed in good condition.Under microscopic inspection, the embolic coil was observed to be damaged with non-concentric loops inside the introducer.A manufacturing record evaluation (mre) was performed for the finished device l11005 number, and no non-conformances related to the reported complaint condition were identified.The device was returned to cerenovus for further evaluation.Visual inspection of the returned device revealed that the core wire was observed in good condition, under microscopic inspection the embolic coil was observed to be damaged with non-concentric loops inside the introducer.The concomitant microcatheter (echelon14 (medtronic's catheter) was not facilitated to the lab to perform further evaluation.However, the instructions for use (ifu) for galaxy g3 indicates that ¿the galaxy g3 microcoil systems compatible with microcatheters with inner lumen diameters ranging from 0.0165 to 0.021 in (0.419 to 0.533 mm)¿.Coil damage is a known potential issue associated with the use of this device.The instructions for use (ifu) provides proper handling instructions for the device to prevent such issue from occurring.The impeded condition could not be tested due to the damage noted on the embolic coil, however, this could be secondary to the issue encountered during the procedure.Therefore, the customer complaint was confirmed.A manufacturing record evaluation (mre) was performed, and no non-conformance was found during the review.It should be noted that product failure could be caused by multiple factors.The instructions for use (ifu) contain the following precautions: do not fasten the rhv valve too tightly around the introducer sheath since excessive pressure may cause damage to the introducer sheath and/or the micro coil as it is advanced into the infusion microcatheter.Additionally, if the introducer tip and microcatheter hub are misaligned, damage may occur to the micro-coil as it passes through this transition.If unusual friction is noticed during advancement or retraction of the micro coil system, verify the locking mechanism or clear tab is unlocked and pulled out from the re-sheathing tool approximately 1 in (2-3 cm).If unusual friction is noticed during advancement or retraction of the micro coil system through the introducer, open the rhv main valve, and partially withdraw the distal end of the introducer to expose its tip within the rhv.Tighten the rhv main valve and flush the y-connector of the rhv with sterile saline and verify that fluid exits the slit in the clear portion of the introducer.After flushing, reinsert the introducer into the infusion catheter hub as described in the ¿microcoil placement¿ section above.If unusual friction is still noticed during advancement or retraction of the microcoil system, verify flush lines are open and properly pressurized.Then slowly withdraw the entire microcoil system and examine for damage.Replace it with a new microcoil system.If friction still exists, withdraw and examine the delivery catheter system.Assignment of root cause for the event remains speculative and inconclusive, based on the limited information provided and the evidence presented by the returned device; however, it is possible that clinical and procedural factors, including device manipulation and device interaction, may have contributed to the reported failures and damages on the returned system.As part of the cerenovus quality process, all devices are manufactured, inspected, and released to approved specifications.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.
|