OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problem Excess Flow or Over-Infusion (1311)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/09/2022
Event Type  malfunction  
Event Description
The customer reported that the subject device would not stop insufflation automatically, so the measuring pressure was higher than the set pressure.The intended procedure was completed without issue.There was no patient harm or user injury reported due to the event.Additional details were requested regarding the reported event; however, no further information was available.
 
Manufacturer Narrative
The customer's report was not confirmed by olympus repair center.It is likely the device could not stop its insufflation automatically because the measuring pressure was higher than the set pressure.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the approved final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, and since no device malfunction was confirmed during evaluation, the definitive root cause of the reported issue could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14043270
MDR Text Key288795773
Report Number8010047-2022-05887
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/05/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-