Model Number 2217-50-041 |
Device Problems
Break (1069); Entrapment of Device (1212)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/22/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
The product experienced problems during procedure.No additional information available.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary: the device associated with this report was returned for analysis.Examination of the physical device can confirm the breakage condition.The end cap of the pinnacle impactor is broken, all pieces were return for investigation therefore the foreign body left in the patient cannot be confirmed.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
|
Event Description
|
Additional information received indicated that the fragment generated remained in the patient.The part broke during the procedure but there was no delay in surgery.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|