| Model Number |
2549-00-010 |
| Medical Device Problem Code |
Contamination /Decontamination Problem (2895)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Instrument tray sent to the hospital dirty still had blood on them from previous hospital and usage.Discovery made during sterile processing/field inspection.No adverse affects on patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : the device associated with this report was not returned for analysis.After photographic examination of the attached evidence, allegation was confirmed.There can be observed that instruments of the reported case presented a foreign substance throughout the surface.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : a manufacturing records evaluation (mre) was not performed as no lot number was provided for this device.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: visual analysis of the returned sample exhibited a 254600501/atune rev stem trial drivr bit presented a foreign substance which seems to be corrosion throughout the surface.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: a manufacturing records evaluation (mre) was not performed as no lot number was provided for this device.,the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Additional information was received indicating that "the ip has been updated to stem trial driver bit" this product was noted to "have potential to collect blood in future cases".For this updated product family, depuy synthes joint reconstruction does not consider this failure to be a reportable malfunction at this time.Further updates will only be provided if additional information is received that changes the regulatory determination.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d4 (lot) and h4.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: d1, d2, d2b, d4 (procode,udi).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned for analysis.After photographic examination of the attached evidence, allegation was confirmed.There can be observed that instruments of the reported case presented a foreign substance throughout the surface.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: a manufacturing records evaluation (mre) was not performed as no lot number was provided for this device.
|
| |
|