Patient allegedly received a gunther tulip implant on (b)(6) 2012 via the right femoral vein due to deep vein thrombosis (dvt) and pulmonary embolism (pe).Patient is alleging tilt.The patient further alleges pain, abdominal pain with movement, and anxiety.Report from ct (computed tomography): "caval perforation: no.Relation of the caval filter to the ivc wall is grade 1 consistent with tenting.Tilt: yes.Right-sided tilt measuring 8.95 degrees.Posterior tilt measuring 3.69 degrees.Migration: apex of the caval filter is at the origin of the right renal vein.".
|
|
The previous medwatch report was submitted by william cook europe under manufacturer report reference# 3002808486-2022-00103.Additional information provided determined that this device was manufactured by cook inc.With the submission of this initial medwatch report, cook inc.Informs that all future submissions regarding this complaint will be handled under the manufacturer report reference # of this initial medwatch report.Initial reporter occupation: non-healthcare professional.Investigation: investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: tilt, pain/abdominal pain, anxiety, physical limitations.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported pain/abdominal pain, anxiety, physical limitations is directly related to the filter and unable to identify a corresponding failure mode at this point in time.A total of 20 devices were manufactured in the reported lot.To date, no other complaints have been reported against the lot.The associated work order was reviewed.No related/relevant notes were documented.The device is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications, and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information become available.
|