| Model Number |
N/A |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of refx4333 showed two other similar product complaint(s) from this lot number.The complaints for this lot number (refx4333) have been reported from the same facility.
|
| |
|
Event or Problem Description
|
|
It was reported "we have a picc that we had to exchange because the purple lumen was leaking at the hub when a needleless connector was attached.We tried both baxter and rhymed needleless connectors and no matter how tight we screwed on the needleless connector it was still leaking.(no issues with red lumen) when needleless connector was removed and syringe attached directly to the line there was no leaking.Leaks from hub connection to needleless connector." additional information received 03/24/2022 a detailed description of the event: purple lumen leaking.Vat tightened connector and leaking stopped.What was infusion when the leak was discovered? cpn any patient symptoms related to fluid leak? no was the catheter removed? catheter removed (b)(6)therapy completed was a new catheter placed? no what type of catheter securement was utilized? securacath was there any patient harm reported? no.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a leaking catheter was confirmed and the cause was manufacturing-related.The product returned for evaluation was one 5fr d/l powerpicc catheter.Usage residues were observed throughout the sample.An attempt to infuse water through the sample using a 12ml syringe revealed both lumens to be patent to infusion; however, leakage was observed at the junction between the purple luer adapter and the syringe.Inspection of the purple luer adapter using an iso taper gauge revealed that the gauge could not be fully inserted into the luer orifice.Microscopic inspection of the purple luer adapter revealed a protrusion in the luer material within the orifice.The mold defect appeared to prevent firm attachment of the cap, as well as full insertion of the taper gauge.The molding defect observed within the luer orifice prevented firm attachment with the syringe and resulted in leakage at the interface between the catheter and the cap.Photographs of the sample have been forwarded to the manufacturing site for further evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported "we have a picc that we had to exchange because the purple lumen was leaking at the hub when a needleless connector was attached.We tried both baxter and rymed needleless connectors and no matter how tight we screwed on the needleless connector it was still leaking.(no issues with red lumen) when needleless connector was removed and syringe attached directly to the line there was no leaking.Leaks from hub connection to needleless connector." additional information received 03/24/2022: 2.A detailed description of the event: purple lumen leaking.Vat tightened connector and leaking stopped.3.What was infusion when the leak was discovered? cpn.4.Any patient symptoms related to fluid leak? no.5.Was the catheter removed? catheter removed 2/23 therapy completed was a new catheter placed? no.6.What type of catheter securement was utilized? securacath.7.Was there any patient harm reported? no.
|
| |
|