ABBOTT VASCULAR EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM Back to Search Results
Model Number 22438-19
Device Problems Break (1069); Stretched (1601)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/16/2022
Event Type  malfunction  
Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that during unpacking of the emboshield nav6 embolic protection system (eps), it was noted that the tip of the barewire was unraveled.Another same size barewire was used to complete the procedure.There was no device use or patient involvement and there was no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
A visual inspection was performed on the returned device.The reported tip break was unable to be confirmed.The reported stretched tip was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents reported from this lot.The investigation determined the reported break and stretched tip appear to be related to case circumstances of the procedure.It is likely that during preparation, the tip of the barewire was handled causing the coils to stretch at the step causing the appearance of a break.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM
Type of Device
EMBOLIC PROTECTION SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14044860
MDR Text Key288811773
Report Number2024168-2022-03691
Device Sequence Number1
Product Code NTE
UDI-Device Identifier08717648137754
UDI-Public08717648137754
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K141678
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number22438-19
Device Catalogue Number22438-19
Device Lot Number1111561
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/16/2022
Initial Date FDA Received04/07/2022
Supplement Dates Manufacturer Received04/29/2022
Supplement Dates FDA Received05/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/15/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-