|
According to the reporter, during completion of dialysis, when the extracorporeal circuit was given back to the patient, a leak of less than one ml was noted from the venous lumen/extension within 0.5cm from the end of the colored luer.The renal vascular access nurse reviewed the catheter and felt it could be repaired.It was said that the treatment will be continued until the repair kit could be ordered and received.No defects/damages were found on the product where the leak was observed as it was difficult to see (only tiny blood leak was visible).Alcoholic chlorhexidine was the cleaning agent used on the device.No ointments were used.There was no luer adapter issue.The insertion site was cleaned with alcoholic chlorhexidine.It was said that it was not a protocol to move the clamps to a new location after each treatment but they were routinely moved for cleaning and dressings etc.Codan bc1000n, standard bloodlines and syringes were being utilized with the device.There was no blood transfusion required.The patient had no medical intervention/treatment due to the event.The catheter was repaired on (b)(6) 2022 and the treatment was continued until (b)(6) 2022 when a further leak was detected (which was separately reported).There was no reported patient injury.
|
|
Additional information: d9, g1, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that there was a leak on the extension tube.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|