MAUDE Adverse Event Report: BAUSCH + LOMB STELLARIS ULTRASOUND HANDPIECE; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAUSCH + LOMB STELLARIS ULTRASOUND HANDPIECE; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number BL3170
Medical Device Problem Code Particulates (1451)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
The user facility in the (b)(6) reported that during phaco, small white particulate matter was ejected into the anterior chamber and then lodged into the irrigation port of the phaco probe causing the anterior chamber to collapse and stopping the irrigation flow.The particulate was removed from the patient's eye and the surgery was completed.There was no patient impact or clinical consequences.
 
Additional Manufacturer Narrative
The product has been requested but not yet received.A review of the device history record found that the serial number of the handpiece met all physical and functional requirements.At this time, no remedial, corrective, preventive, or field safety corrective action is required.The investigation is ongoing.
 
Additional Manufacturer Narrative
Correction: investigation findings code: removed 3221 and replaced it with 202.
 
Additional Manufacturer Narrative
The handpiece was not returned for evaluation, but the bl5114 disposable pack was returned for evaluation.One small plastic vial containing a light blue irrigation sleeve was received in a plastic bag.The original packaging was not received.The part number and lot number cannot be verified or determined.The irrigation sleeve was very dirty with particulates, saline dendrites, and dried solutions all over it.The tip of the sleeve was inverted.There were white-ish-cream colored particles visible in the sleeve, tip area, and inside the base of the hub area.The particles were removed and placed inside a petri-dish.The sleeve was cleaned and inspected.The tip was torn, and a small piece of the tip material has been torn off and was missing.Three particles were noted and white-ish in color.These particles are hard to the touch but break apart if pressure is applied.One particle did break into two pieces.Therefore, 4 particles are visible.The largest particle is approximately 3900 microns long x 2720 microns wide.These particles were taped to the bottom of the petri-dish and were sent to an outside lab for further analysis.The white particles were analyzed using an energy-dispersive spectrometer (eds) equipped scanning electron microscope (sem).Eds analysis indicated that the white particles consist primarily of calcium.Lesser concentrations of phosphorus, sodium, oxygen, carbon, potassium, chlorine, and magnesium were also detected.This is consistent with calcium phosphate and saline residue.These materials are not used in the bl5114 product and the source of the returned particles cannot be determined.The trend analysis.Risk analysis, and/or directions for use were considered acceptable, with the product performing within anticipated rates.No corrective action required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STELLARIS ULTRASOUND HANDPIECE
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
BAUSCH + LOMB
rochester NY 14609
Manufacturer (Section G)
BAUSCH + LOMB, INC.
3365 tree court industrial blv
st. louis MO 63122
Manufacturer Contact
juli moore
3365 tree court industrial blv
st. louis, MO 63122
MDR Report Key14045753
Report Number0001920664-2022-00045
Device Sequence Number4775599
Product Code HQC
UDI-Device Identifier00757770480852
UDI-Public(01)00757770480852(11)200429
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K063331
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBL3170
Device Catalogue NumberBL3170
Device Lot Number222
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/18/2022
Supplement Date Received by Manufacturer04/28/2022
04/28/2022
Initial Report FDA Received Date04/07/2022
Supplement Report FDA Received Date05/24/2022
06/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/29/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BL5114, STELLARIS SYSTEM, AND ACCESSORIES.
Patient SexUnknown
-
-