• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I 3I T3® EX HEX PARALLEL WALLED IMPLANT 4 X 10MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I 3I T3® EX HEX PARALLEL WALLED IMPLANT 4 X 10MM; DENTAL IMPLANT Back to Search Results
Model Number BOES410
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/12/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that dr.Indicated malfunction.The doctor tried to remove the mount, but it did not come off and the implant body was removed.It has been treated and is not re-embedded.Scheduled to be embedded at a later date.No health hazard.Malfunction, does not come off the mount.Tooth site # 30.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
One (1) 3i t3â® ex hex parallel walled implant 4 x 10mm (boes410) was returned for investigation.Visual evaluation of the as returned product identified damage mount and drive feature of the mount.Unable to separate.Functional testing to recreate the reported event, verified the devices do not disengage.Additionally, the coob could not be verified.Pre-existing conditions noted on the per was moderate bone density (type ii), bruxism, clenching, malocclusion.The reported device was located on tooth # 30 (universal) and was placed and removed on the same day.The customer did not provide any pictures or x-rays.Appropriate documentation was reviewed.Dhr review was completed for the subject lot number (2018070989).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review was performed for the reported lot number (2018070989) for similar events and no other complaint was identified.July post market trending was reviewed and there were no actionable events or corrective actions for the reported event.No actionable items have been triggered that will affect complaint handling on our end for this month.Therefore, based on the available information, device malfunction did occur and the reported event was confirmed.Additionally, the coob could not be verified as the product was used, and the conditions when it was received are unknown / non-verifiable.The following sections have been updated: b4: date of this report d9: device availability g3: date received by manufacturer g6: checked "follow-up" h2: checked follow-up type h3: changed "no" to "yes" h6: entered evaluation codes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3I T3® EX HEX PARALLEL WALLED IMPLANT 4 X 10MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key14046121
Report Number0001038806-2022-00504
Device Sequence Number4895020
Product Code DZE
UDI-Device Identifier00844868032441
UDI-Public(01)00844868032441(17)230723(10)2018070989
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K133049
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 09/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/23/2023
Device Model NumberBOES410
Device Catalogue NumberBOES410
Device Lot Number2018070989
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/30/2022
Supplement Date Received by Manufacturer09/29/2022
Initial Report FDA Received Date04/07/2022
Supplement Report FDA Received Date09/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/23/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age55 YR
Patient SexMale
-
-