• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: K2M, INC. INNER SHAFT; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

K2M, INC. INNER SHAFT; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR Back to Search Results
Model Number 7012-90096
Medical Device Problem Codes Fracture (1260); Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Hospital discarded.
 
Event or Problem Description
It was reported that a mojave inner shaft broke intra-operatively.There were no adverse consequences to the patient.The procedure was completed successfully without surgical delay.
 
Additional Manufacturer Narrative
Visual, dimensional, material and functional analysis could not be performed as the device associated with this record was not returned.Device history records could not be reviewed as a valid lot number was not provided and could not be obtained.Complaint history records were reviewed for this catalog number, no adverse trends were identified.It was reported that a mojave inner shaft broke intra-operatively during cage expansion.The surgery was being performed at l4-s1.The device was not used at a difficult angle and the surgeon did not appear to apply excessive force when the issue occurred.The device has been used 2.5 years multiple times.Since the device was not returned, an exact cause of the reported event could not be determined.Potential causes include: normal wear due to device age/ extensive use, application of excess force during use, tip material wear, etc.
 
Event or Problem Description
It was reported that a mojave inner shaft broke intra-operatively.There were no adverse consequences to the patient.The procedure was completed successfully without surgical delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNER SHAFT
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer (Section G)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key14047017
Report Number3004774118-2022-00132
Device Sequence Number20054782
Product Code MAX
UDI-Device Identifier10888857391703
UDI-Public10888857391703
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K163364
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7012-90096
Device Catalogue Number7012-90096
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/11/2022
Supplement Date Received by Manufacturer05/31/2022
Initial Report FDA Received Date04/07/2022
Supplement Report FDA Received Date06/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-