• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD RESURFACING FEMORAL HEAD 42MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD RESURFACING FEMORAL HEAD 42MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74121142
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Failure of Implant (1924); Inadequate Osseointegration (2646)
Date of Event 03/09/2020
Type of Reportable Event Serious Injury
Event or Problem Description
Us legal.It was reported that the plaintiff underwent a right hip revision surgery on (b)(6) 2020 due to femoral component subsidence.During the revision surgery, only the bhr femoral head was explanted.The patient tolerated the procedure well.The primary bhr surgery was performed on (b)(6) 2010.
 
Additional Manufacturer Narrative
Internal complaint reference: case (b)(4).
 
Additional Manufacturer Narrative
Internal complaint reference number: (b)(4).Section d4 and h4 were corrected.
 
Additional Manufacturer Narrative
H3, h6: it was reported that right hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.A review of the historical complaints data for the femoral head was performed.Using batch numbers to evaluate patterns of repeated failures or defects over the lifetime of the product and using part numbers and the reported failure modes to evaluate patterns of repeated failures or defects in a timeframe prior 12 months as of the complaint aware date.Other complaints were identified to involve this batch.No other similar complaints were identified for the part number and the reported failure mode.However, as the device is no longer sold, no action is to be taken.In the absence of the actual devices, the production records were reviewed for the known devices reportedly involved in this incident.All the released devices involved met manufacturing specifications upon release for distribution.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions required.The available medical documents were reviewed.Without the supporting imaging, complete revision operative report and/or the analysis of the explanted components, the root cause of the reported subsidence cannot be confirmed, and it cannot be concluded that the reported event was associated with a mal performance of the implant.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.Based on the information provided we cannot confirm or further investigate the reported complaint, our investigation remains inconclusive, and a definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.If additional information becomes available in the future, this case will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
 
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).H3, h6: it was reported that right hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.A review of the historical complaints data for the femoral head was performed.Using batch numbers to evaluate patterns of repeated failures or defects over the lifetime of the product and using part numbers and the reported failure modes to evaluate patterns of repeated failures or defects in a timeframe prior 12 months as of the complaint aware date.Other complaints were identified to involve this batch.No other similar complaints were identified for the part number and the reported failure mode.However, as the device is no longer sold, no action is to be taken.In the absence of the actual devices, the production records were reviewed for the known devices reportedly involved in this incident.All the released devices involved met manufacturing specifications upon release for distribution.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions required.The available medical documents were reviewed.Without the supporting imaging, complete revision operative report and/or the analysis of the explanted components, the root cause of the reported subsidence cannot be confirmed, and it cannot be concluded that the reported event was associated with a mal performance of the implant.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.Based on the information provided we cannot confirm or further investigate the reported complaint, our investigation remains inconclusive, and a definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.If additional information becomes available in the future, this case will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESURFACING FEMORAL HEAD 42MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14047232
Report Number3005975929-2022-00130
Device Sequence Number17359745
Product Code NXT
UDI-Device Identifier03596010502773
UDI-Public3596010502773
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P040033
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2020
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 08/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2013
Device Catalogue Number74121142
Device Lot Number085299
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/15/2022
Supplement Date Received by Manufacturer07/08/2022
07/29/2022
08/04/2022
Initial Report FDA Received Date04/07/2022
Supplement Report FDA Received Date07/12/2022
07/31/2022
08/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/03/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ACETLR CUP HAP 48MM W/ IMPTR-090247
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age64 YR
Patient SexFemale
-
-