Event or Problem Description
It was reported that a patient underwent a breast mass excision procedure on an unknown date and suture was used.During use, while sewing the skin, the suture pulled off the needle.Changed another one to complete the surgery.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: was the needle piece removed from the patient? unknown.Patient condition? the patient was in good condition.This is a combination product, and the event has been reviewed for both the suture and the triclosan.Trade name - irgacare®.Active ingredient(s) ¿ triclosan.Dosage form ¿ suture/solid/parenteral.Strength ¿ = 275 g/m.This report is being submitted pursuant to the provisions of 21 cfr part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 8/9/2022.This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that an opened sample of product code vcp441h was received for evaluation.The visual inspection concluded that in the detached needle, the swage and attachment area was noted to be as expected and no suture remnant could be observed at the barrel hole, and the shape of the barrel shows there isn¿t enough impression, resulting in a needle pull off defect.Based on the information currently available, the pull-out was identified during the investigation of the sample received.This defect is caused when does not tighten the press correctly and the swage area is weak on the needle/suture this product issue will be addressed through ethicon quality system.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Trade name - irgacare®; active ingredient(s) ¿ triclosan; dosage form ¿ suture/solid/parenteral; strength ¿ = 275 g/m.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL+ UD 27IN 4-0 S/A FS-1
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC. - BRAZIL
rodovia presidente
dutra km 154
sao paolo 12240 -908
BR   12240-908
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
MDR Report Key14047405
Report Number2210968-2022-02490
Device Sequence Number4791894
Product Code GAM
UDI-Device Identifier10705031053274
UDI-Public10705031053274
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
K032420
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Model NumberVCP441H
Device Catalogue NumberVCP441H
Device Lot NumberPJ4659
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/15/2022
Supplement Date Received by Manufacturer07/26/2022
Initial Report FDA Received Date04/07/2022
Supplement Report FDA Received Date08/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/09/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexFemale
-
-