| Model Number |
B2050-060 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported balloon rupture appears to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a 98% stenosed lesion in the left superficial femoral artery (sfa) with heavy calcification and moderate tortuosity.A 5.00 x 60mm armada balloon dilatation catheter (bdc) was advanced to the target lesion on a 0.035 unspecified guidewire; however, during the second inflation the balloon ruptured at 8 atmospheres after 2 minutes due to heavy calcification in the lesion.The bdc was removed from the patient and a 6.00 x 80mm absolute stent was implanted.Then post-dilatation was performed with a 5.00 x 40mm armada balloon to complete the procedure.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information regarding this issue was provided.
|
| |
|