• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION NC QUANTUM APEX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION NC QUANTUM APEX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 6534
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in the moderately calcified coronary artery.A 30mm x 2.50mm nc quantum apex balloon catheter was advanced for dilation.However, during inflation at 10 atmospheres, the balloon ruptured.The procedure was completed with a different device and there were no patient complications nor injuries reported.
 
Additional Manufacturer Narrative
Date of event: used the first date of the month of the aware date as no event date was provided.
 
Additional Manufacturer Narrative
B3: date of event: used the first date of the month of the aware date as no event date was provided.Device evaluated by mfr.: returned product consisted of an nc quantum apex balloon catheter.The device was visually and microscopically examined.There was contrast in the inflation lumen, and the balloon was loosely folded.At 16mm from the tip moving proximally for a length of 16mm, the balloon has a longitudinal tear.Product analysis confirmed the reported burst, as there was a longitudinal tear in the balloon.
 
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in the moderately calcified coronary artery.A 30mm x 2.50mm nc quantum apex balloon catheter was advanced for dilation.However, during inflation at 10 atmospheres, the balloon ruptured.The procedure was completed with a different device and there were no patient complications nor injuries reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC QUANTUM APEX
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14048311
Report Number2134265-2022-04154
Device Sequence Number20303734
Product Code LOX
UDI-Device Identifier08714729783572
UDI-Public08714729783572
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K121667
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/20/2024
Device Model Number6534
Device Catalogue Number6534
Device Lot Number0028242424
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/01/2022
Supplement Date Received by Manufacturer09/16/2022
Initial Report FDA Received Date04/07/2022
Supplement Report FDA Received Date09/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-