| Model Number |
BIOCON-900 |
| Medical Device Problem Codes |
Smoking (1585); Sparking (2595)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No product was returned for evaluation.Should new relevant information become available, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the staff plugged in device and they said it spared and looks like there was smoke.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that staff plugged in the device and they said it sparked and looks like there was smoke.And it was also reported that there is no patient involved in the event and the power cord and inlet of the device were replaced by a field service engineer.Mcube technology investigated the power cord and inlet and found that among the 3 pins of the inlet, the ground pin was damaged, and it was inserted into the plug.The cause of the event could not be confirmed.The power cord and inlet were ul approved and the electrical safety of the device complies with iec60601-1.
|
| |
|
Event or Problem Description
|
|
It was reported that the staff plugged in device and they said it sparked and looks like there was smoke.
|
| |
|