AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number PCB00
Medical Device Problem Codes Crack (1135); Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the preloaded intraocular lens (iol) had deploying issue.When attempting to deploy the lens, the injector bulged outwards on the side and the lens was stuck very close to the tip where it bulged.There was contact with patient's right eye.There were no medical or surgical interventions required.Another lens of same model and diopter was used as the replacement.The account used ophthalmic viscoelastic device (ovd) which was allowed to acclimatize fully to operating room (or) temperature.After inserter prep, there was a delay of 2-3 minutes in the surgery before an attempt was made to advance the lens.The lens was not fully deployed until handed off to doctor.The lens initially moved a little, then was stuck which reportedly looked like there was a defect with cartridge itself.No further information was provided.
 
Additional Manufacturer Narrative
If implanted, give date: not applicable, as there is no indication that the lens was implanted if explanted, give date: not applicable, as there is no indication that the lens was implanted; therefore, not explanted.The device was not returned for analysis.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Section d9: device available for evaluation: yes.Section d9: returned to manufacturer on: apr 22, 2022.Section h3: device evaluated by manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection under magnification revealed that the handpiece was received with the lens stuck inside of the cartridge and that the lens was overridden by the plunger rod.The cartridge and cartridge tip were both observed to have a bulge where the lens was stuck.The handpiece was disassembled and while no assembly issues were identified, the plunger rod was observed to be bent.The lens was removed from the cartridge and was observed to be damaged.Conclusion: the reported complaint issues were observed during product evaluation; however, based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.Therefore, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key14048426
Report Number3012236936-2022-00932
Device Sequence Number14841602
Product Code HQL
UDI-Device Identifier05050474558212
UDI-Public(01)05050474558212(17)220716
Combination Product (Y/N)N
Initial Reporter StateWY
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/16/2022
Device Model NumberPCB00
Device Catalogue NumberPCB0000175
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/15/2022
Supplement Date Received by Manufacturer06/21/2022
Initial Report FDA Received Date04/07/2022
Supplement Report FDA Received Date06/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/16/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
Patient Weight59 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-