MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Model Number |
3545455 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Codes |
Swollen Lymph Nodes/Glands (4432); Breast Discomfort/Pain (4504)
|
| Date of Event |
03/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: material rupture, breast pain, lymphadenopathy.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) caucasian female patient underwent primary breast augmentation with two 455cc mentor memorygel breast implants.Post-operatively, the patient was diagnosed, via physical examination and mammogram, with right breast implant rupture and silicone outside the breast tissue.The right breast also had pain.As a result, the patient underwent breast implant removal and replacement surgery on (b)(6) 2022.During the revision surgery, it was discovered that the left implant was also ruptured.Subsequently, both implants were removed and replaced with two 480cc mentor ultra high smooth implants.This medwatch form is for the right breast prosthesis.
|
| |
|
Additional Manufacturer Narrative
|
|
On april 19, 2022, the mentor failure analysis lab received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
On may 4, 2022, the product investigation was completed.Device investigation summary: the product was returned to mentor for evaluation.Mentor then conducted visual inspection and microscopic examination of the returned device.Visual analysis of the returned sample revealed that the siltex uhp 455cc breast implant had a tear at the juncture on the posterior patch.Additionally, parallel lines of shell wear were observed on the anterior aspect.Microscopic examination was performed and the cause of the tear could not be identified.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Although no conclusion could be reach on the cause of the reported event, the instructions for use contain the following caution: rupture can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.The following things may cause implant to rupture: damage by surgical instruments; stressing the implant during implantation and weakening it; folding or wrinkling of the implant shell; excessive force to the chest (e.G.During closed capsulotomy); trauma; compression during mammographic imaging; and severe capsular contracture.Breast implants may also simply wear out over time.Most women undergoing augmentation or reconstruction with a mammary prosthesis will experience some pain postoperatively.While pain normally subsides in most women as they heal from surgery, it can become a chronic problem in other women.Chronic pain can be associated with a variety of factors.Surgeons should instruct their patients to inform them if there is significant pain or if pain persists.Pain is a known complication associated with these devices and is referenced in our current product insert data sheet.Lymphadenopathy or palpable lymph nodes may be the result of bacterial or viral infection, neoplasm (primary or metastatic), or as the result of an idiopathic inflammatory process or foreign body reaction.There have been reports regarding movements of silicone gel material to lymph nodes even with or without evident of rupture leading to lymphadenopathy.A number of epidemiology studies have evaluated large populations of women with breast implants from a variety of manufacturers and implant models.The studies do not, taken together, support an association of breast implants and a diagnosed rheumatic disease.Other than one study, these studies do not distinguish whether the women had ruptured or intact implants.Because there were no supporting pathology reports or physician consultation summaries forwarded to product evaluation, we are unable to confirm the presence of or the possible etiology of the reported lymphadenopathy.Lymphadenopathy is a known complication associated with this surgery and is referenced in our current product insert data sheet.As part of mentor¿s quality process all devices are manufactured, inspected, and released to approved specifications.
|
| |
|