MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EASYTRAK 2; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EASYTRAK 2; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D Back to Search Results
Model Number 4517
Medical Device Problem Codes Failure to Capture (1081); Fracture (1260); High impedance (1291); High Capture Threshold (3266)
Health Effect - Clinical Codes Dyspnea (1816); Swelling/ Edema (4577)
Date of Event 02/02/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient attended clinic for a routine follow-up appointment, and indicated hearing beeping tones coming from the related pulse generator (pg).During device interrogation, it was found that the pg had reached elective replacement indicator (eri) as expected.Unexpectedly, it appeared that this this left ventricular (lv) lead had a suspected fracture because it was not capturing on presentation.The patient also indicated that he was experiencing shortness of breath, and that his legs had become more swollen.The lead trend graph showed a change in impedance measurements, which were high and out of range.The lv lead configurations were tested, confirming that the impedances were greater than 2500 ohms, and high capture thresholds were also observed.Since the patient had good underlying rhythm, the physician elected to reprogram device settings.The patient has been scheduled for routine replacement of the related pg; therefore, it is intended that this lead will be replaced at the same time.This lead remains implanted and in service.No adverse patient effects were reported.
 
Event or Problem Description
It was reported that the patient attended clinic for a routine follow-up appointment, and indicated hearing beeping tones coming from the related pulse generator (pg).During device interrogation, it was found that the pg had reached elective replacement indicator (eri) as expected.Unexpectedly, it appeared that this this left ventricular (lv) lead had a suspected fracture because it was not capturing on presentation.The patient also indicated that he was experiencing shortness of breath, and that his legs had become more swollen.The lead trend graph showed a change in impedance measurements, which were high and out of range.The lv lead configurations were tested, confirming that the impedances were greater than 2500 ohms, and high capture thresholds were also observed.Since the patient had good underlying rhythm, the physician elected to reprogram device settings.The patient has been scheduled for routine replacement of the related pg; therefore, it is intended that this lead will be replaced at the same time.This lead remains implanted and in service.No adverse patient effects were reported.A follow-up email was sent to the field to obtain further information regarding the status of the lead.No further information was able to be provided.
 
Additional Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYTRAK 2
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key14048940
Report Number2124215-2022-03753
Device Sequence Number5995092
Product Code NIK
UDI-Device Identifier00802526417665
UDI-Public00802526417665
Combination Product (Y/N)Y
Initial Reporter CountryAS
PMA/510(K) Number
P010012/S024
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2006
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/21/2008
Device Model Number4517
Device Catalogue Number4517
Device Lot Number424711
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/02/2022
Supplement Date Received by Manufacturer04/26/2022
Initial Report FDA Received Date04/07/2022
Supplement Report FDA Received Date05/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/21/2006
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age48 YR
Patient SexUnknown
-
-