| Model Number |
AORTIC CUTTERS, 5-PACK (4.3MM) |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Trackwise id (b)(4).Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.
|
| |
|
Event or Problem Description
|
|
The hospital reported that during a coronary artery bypass procedure the aortic cutters, 5-pack (4.3mm) unlock button was harder than usual.They didn't use it just in case, but they opened a new product of the same size and could use it without any problem.There is no damage to the patient's health.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
Trackwise # (b)(4).Analysis of production: (3331/213/67) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/213/67) the review of the historical data indicates that no other similar complaints was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110/213/67) the overall 24 month product complaint trend data for the period apr 2020 through mar 2022 was reviewed.There were no triggers identified for the review period.
|
| |
|