MAUDE Adverse Event Report: RESPIRONICS, INC. TRILOGY 100; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. TRILOGY 100; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 1054260
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging visualization of particles in the air path.There was no report of patient harm or injury.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Additional Manufacturer Narrative
The manufacturer previously reported receiving information alleging visualization of particles in the air path.There was no report of serious patient harm or injury.Repeated attempts to have the device and components returned for evaluation and investigation were unsuccessful.The manufacturer believes they will be unable to gather additional information.The manufacturer is submitting a final report at this time.If pertinent information becomes available to the manufacturer at a later date, an addendum to this final report will be filed.
 
Additional Manufacturer Narrative
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously reported receiving information alleging visualization of particles in the air path.There was no report of serious patient harm or injury.The device was returned to the manufacturer's service center for further evaluation.The device was evaluated.There was rear case enclosure damaged.The internal aspect of the device was inspected.The device powered on and airflow was confirmed.The device's downloaded logs were reviewed by the manufacturer.There were no errors found.The manufacturer concludes that they could not confirm the customer's allegation and there was no particles verified in airpath and unit is scrapped.In this report, sections d9, g3, h2, h3, and h6 have been updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 100
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
rosko melissa
6501 living place
pittsburgh, PA 15206
7243512041
MDR Report Key14052084
Report Number2518422-2022-12769
Device Sequence Number4923467
Product Code CBK
UDI-Device Identifier00606959026308
UDI-Public00606959026308
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K083526
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1054260
Device Catalogue Number1054260
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/11/2022
Supplement Date Received by Manufacturer07/06/2023
02/05/2025
Initial Report FDA Received Date04/08/2022
Supplement Report FDA Received Date07/10/2023
03/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/04/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
FDA RES 88071
Patient Sequence Number1
Patient SexUnknown
-
-