ZIMMER BIOMET, INC. HFN RH 130 DEG 9MM X 380MM; ROD, FIXATION Back to Search Results
Model Number 8145-09-380
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient had affixus nail implanted.Subsequently, revision procedure was performed to remove the fractured nail and external fixation was put in place.
 
Additional Manufacturer Narrative
This event was initially reported under mfr# : 0009613350 -2021 -00301.Event date: unknown day in (b)(6) 2021.Implant date: unknown day in (b)(6) 2020.Concomitant products: cat: 814510100 lot: wg1111300a hfn lag screw 10.5mm x 100mm.Cat: unknown screw lot: unknown.Cat: unknown screw lot: unknown.Report source: foreign- (b)(6).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2022 - 00848, 0001825034 - 2022 - 00849, 0001825034 - 2022 - 00850.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d4, g3, g6, h1, h2, h3, h4, h6, h10 complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned product identified that the nail is fractured.Fracture analysis was done on the nail and fatigue fractures were identified.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HFN RH 130 DEG 9MM X 380MM
Common Device Name
ROD, FIXATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key14052340
Report Number0001825034-2022-00851
Device Sequence Number14695944
Product Code HSB
UDI-Device Identifier00887868038914
UDI-Public(01)00887868038914(17)250531(10)973540
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K100238
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Expiration Date05/31/2025
Device Model Number8145-09-380
Device Catalogue Number814509380
Device Lot Number973540
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/11/2022
Supplement Date Received by Manufacturer06/03/2022
Initial Report FDA Received Date04/08/2022
Supplement Report FDA Received Date06/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/19/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
-
-