| Model Number |
4540008-02 |
| Medical Device Problem Code |
Excess Flow or Over-Infusion (1311)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported by the user facility / translation of user facility information by bbm sales organization in (b)(6): "overinfusion" "reason of complaint: we've had another easypump returned to us yesterday where it was reported that it infused too quickly.Patient was connected at approx.1600 however easypump was found to be empty by the nurse at 1200pm next day.Patient's wife then reported that it was already empty at 0700.".
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun melsungen ag internal report (b)(4) the complaint is still under investigation.A follow-up report will be provided, as soon as investigation has been completed.Note: this report is being filed for an item number that is not sold in the united states, however similar items are sold in the united states by b.Braun medical, inc.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun melsungen ag internal report # (b)(4).Received 1 used and filled easypump ii lt 270-27-s.As received condition, clamp clip of received sample was clamped.Patient connector of received sample was not connected to wing cap.Visual inspection had done throughout the sample.According to customer, the pump was emptied.However, during investigation, it was observed that there is remaining solution inside the easypump.There was still approximately 15ml of solution inside the pump.The sample was sent for flow rate test.The flow rate deviation from nominal flow rate for the received samples were -5.57% from nominal flow rate.The flow rate of received samples were within the specification ±15% deviation from nominal flow rate.Summary of root cause analysis: the sample was not emptied when received.Furthermore, the flow rate of received sample was within the specification, hence we considered this complaint as not confirmed.Received sample is within specification.Device history record (dhr): reviewed the dhr for batch 21d28ge221, there is no abnormality and no such defect detected at in process and at final control inspection.
|
| |
|
Search Alerts/Recalls
|