| Model Number |
8884720825E |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Needle Stick/Puncture (2462)
|
| Date of Event |
03/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Customer reports: when attempting to remove the stylet, the green cap became disconnected from it.The nurse's skin was punctured by the exposed wire that remained in the tube.According to the nurse, the tube was flushed prior to attempting to remove the stylet.
|
| |
|
Additional Manufacturer Narrative
|
|
Other selected for no sample return because the product was discarded.An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record (dhr) was reviewed, and no abnormal process conditions were present during the manufacturing of the product that could have led to the reported defect from customer complaint.The dhr review showed that all acceptance criteria inspections per established sampling levels were within acceptable limits during the production process.The product sample was not provided for evaluation therefore the reported condition could not be confirmed.The root cause and corrective actions could not be identified without a sample evaluation however, the potential cause of the reported condition most likely occurred if there was a failure to activate the hydromer making it difficult to remove the stylet from the feeding tube.Please refer to the instructions for use (ifu) for the proper procedure for insertion of the feed tube and extraction of the stylet which reads: "using a syringe, inject approximately 10 cc of water into the tube to activate the hydromer coating.¿ based on the analysis, the observed occurrence of harm (o) is (1, improbable) which is equal the range of the expected occurrence of harm (o) (4, improbable).
|
| |
|
Search Alerts/Recalls
|