MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number FX-410360
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Seroma (2069); Lymphoma (3263)
Date of Event 02/28/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported left side "exchange of textured breast implants due to the patient¿s concern with the product" and "fluid build up in the left side cavity per mri and swelling." healthcare professional additionally reported "positive alcl diagnosis." cytology report provided "the tumor cells are diffusely positive for cd45, cd3, cd30, and gata3, and negative for ae1/ae3 and ck7." pathological marker confirming alk- has not been received.Device has been explanted.
 
Additional Manufacturer Narrative
(b)(4).A review of the device history record has been completed.No deviations or non-conformances noted.The events of seroma and lymphoma-alcl-suspected are physiological complications and analysis of the device generally does not assist allergan in determining a probable cause for these events.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: "exchange of textured breast implants due to the patient¿s concern with the product," "fluid build up in the left side cavity per mri and swelling," and "positive alcl diagnosis.".
 
Additional Manufacturer Narrative
Additional healthcare professional contact information: oncologist dr.(b)(6), phone number: (b)(6).Device evaluation: visual analysis of the returned device identified: weight to the spec, opening, crease fold, wear abrasion.A microscopic analysis was performed which identify a striated edge opening.Based on the device analysis the final assessment is: a striated edge opening on radius side assessed as surgical damage.
 
Event or Problem Description
Healthcare professional reported left side "exchange of textured breast implants due to the patient¿s concern with the product" and "fluid build up in the left side cavity per mri and swelling." healthcare professional additionally reported "positive alcl diagnosis." cytology report provided "the tumor cells are diffusely positive for cd45, cd3, cd30, and gata3, and negative for ae1/ae3 and ck7." pathological marker confirming alk- has not been received.Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
chris sprague
12331-a riata trace parkway
building 3
austin, TX 78727
8479375615
MDR Report Key14053499
Report Number9617229-2022-05627
Device Sequence Number4766213
Product Code FTR
UDI-Device Identifier10888628002975
UDI-Public10888628002975
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P040046
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/07/2022
Device Model NumberFX-410360
Device Catalogue NumberFX-410360
Device Lot Number3019120
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/01/2022
Supplement Date Received by Manufacturer04/29/2022
Initial Report FDA Received Date04/08/2022
Supplement Report FDA Received Date05/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/07/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
Patient SexFemale
Patient Weight108 KG
-
-