C.R. BARD, INC. (BASD) -3006260740 DOT PG PRO 18G 10CM FULL; CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Codes Break (1069); Material Frayed (1262)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 03/12/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of refy1539 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported "at approximately 10:30 am today, (b)(6) 2022, there was an equipment issue with a midline device.Md attempted to insert a midline into patient.The device malfunctioned, dr called surgery to aid in the removal of the midline from patient, which was successful.No injury to patient; patient denies pain; va duplex of the affected limb ordered.We recovered the device.".
 
Event or Problem Description
It was reported "at approximately 10:30 am today, (b)(6) 2022, there was an equipment issue with a midline device.Md attempted to insert a midline into patient.The device malfunctioned, dr called surgery to aid in the removal of the midline from patient, which was successful.No injury to patient; patient denies pain; va duplex of the affected limb ordered.We recovered the device.".
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and batch history, applicable previous investigation(s), applicable manufacturing records, sample analysis and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of an equipment issue with a powerglide pro was confirmed but the cause is unknown.Three photographs were returned for evaluation.The 18g x 10cm powerglide pro contained residual material in the lumen of the catheter, indicating the device had been used.A guidewire fragment was visible in the photographs.The fragment contained a complete break in the inner core and outer coil wires.The coil wire on the guidewire was unraveled and elongated.It is possible that complications with device deployment contributed to the guidewire damage.Damage to the guidewire can occur if the guidewire is forced against the sharp edge of the needle bevel causing it to shear.However, without the physical sample, the device could not be microscopically examined, tactility evaluated, or dimensionally analyzed.Therefore, the exact root cause could not be determined.H3 other text: evaluation findings are in section h11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOT PG PRO 18G 10CM FULL
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
johanna de oliveira
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key14053568
Report Number3006260740-2022-01201
Device Sequence Number11913013
Product Code FOZ
UDI-Device Identifier00801741110665
UDI-Public(01)00801741110665
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K153280
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberF218101
Device Lot NumberREFY1539
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/21/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/15/2022
Supplement Date Received by Manufacturer05/06/2022
Initial Report FDA Received Date04/08/2022
Supplement Report FDA Received Date05/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-