C.R. BARD, INC. (BASD) -3006260740 POWERPICC PROVENA SOLO 5F TL MAX BARRIER BARRIER PLUS KIT W/BIOPATCH; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of refx0108 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported "stylet checked prior to insertion.Picc inserted in pt.'s left arm, some minor insertion difficulty noted advancing past subclavian area, resolved easily with repositioning pt.'s left arm.Stylet checked upon removal from picc after picc successfully inserted and kink noted within last 2cm of stylet (recognized visually and tactilely).".
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and lot history review, applicable previous investigation(s), sample (if available), applicable manufacture records, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a kinked stylet is confirmed; however, the exact cause is unknown.One 3cg stylet was returned for evaluation.An initial visual observation showed no blood residue on the returned sample.Bends were observed in the returned stylet approximately 1 cm and 3.6 cm proximal to its distal tip as well as approximately 20.2 cm distal to the yellow handle of the stylet.No breaks were observed in the core wire of the stylet and the distal tip was found to be present and intact.A microscopic observation revealed no breaks or tears in the tubing at the distal end of the stylet, and the slight bends within this tubing were found to be near the interfacing of two magnets.Insufficient evidence was found on the returned sample to indicate a specific root cause of the bends in the returned stylet; however, possible contributing factors include advancement of the stylet against resistance and damage during manipulation, handling, or use.A lot history review (lhr) of refx0106 showed two other similar product complaint(s) from this lot number.H3 other text : evaluation findings are in section h.11.
 
Event or Problem Description
It was reported "stylet checked prior to insertion.Picc inserted in pt.'s left arm, some minor insertion difficulty noted advancing past subclavian area, resolved easily with repositioning pt.'s left arm.Stylet checked upon removal from picc after picc successfully inserted and kink noted within last 2cm of stylet (recognized visually and tactilely).".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC PROVENA SOLO 5F TL MAX BARRIER BARRIER PLUS KIT W/BIOPATCH
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
johanna de oliveira
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key14053573
Report Number3006260740-2022-01202
Device Sequence Number14603281
Product Code LJS
UDI-Device Identifier00801741155123
UDI-Public(01)00801741155123
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K072230
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberS1395108D4
Device Lot NumberREFX0106
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/04/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/15/2022
Supplement Date Received by Manufacturer06/06/2022
Initial Report FDA Received Date04/08/2022
Supplement Report FDA Received Date06/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-