MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET XXL 360 12MM X 45CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET XXL 360 12MM X 45CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0036181240
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/05/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the procedure the coil (subject device) was prematurely detached/separated inside the microcatheter before energizing the inzone.The physician replaced it with a new device and continued the procedure without clinical consequences to the patient.
 
Additional Manufacturer Narrative
The device is not available to manufacturer.
 
Additional Manufacturer Narrative
Due to the automated mes (manufacturing execution system) system there are controls in the manufacturing process to ensure the product met specifications upon release.During visual inspection, the target device was returned together with unknown/non-stryker microcatheter and balloon catheter, the coil was found to be stuck inside the balloon catheter, the coil delivery wire was found to be severely kinked/bent, the main coil was found to be detached, the main coil was found to be stretched, the suture was found to be intact, and significant amount of blood and blood clots were noted within the catheter and on the coil.A functional test was not available.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint was confirmed based on analysis.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.During analysis, the main coil was found to be stretched and detached from the delivery wire and stuck inside the microcatheter due to significant amount of blood and blood clots.An assignable cause of procedural factors will be assigned to the as reported/as analyzed 'main coil prematurely detached/separated during use' and to the analyzed ¿main coil stretched', as these defects appear to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu but due to procedural and/or anatomical factors during use, the product performance was limited.An assignable cause of handling damage will be assigned to the as analyzed 'coil delivery wire kinked/bent' since this issue is associated with handling of the device during the clinical procedure.
 
Event or Problem Description
It was reported that during the procedure the coil (subject device) was prematurely detached/separated inside the microcatheter before energizing the inzone.The physician replaced it with a new device and continued the procedure without clinical consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET XXL 360 12MM X 45CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key14053918
Report Number3008881809-2022-00162
Device Sequence Number4821447
Product Code HCG
UDI-Device Identifier07613327128307
UDI-Public07613327128307
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K161429
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/03/2022
Device Model NumberM0036181240
Device Catalogue NumberM0036181240
Device Lot Number21903235
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/06/2022
Supplement Date Received by Manufacturer06/03/2022
Initial Report FDA Received Date04/08/2022
Supplement Report FDA Received Date06/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/05/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-