• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPICC CATHETER WITH SHERLOCK 3CG TIP POSITIONING SYSTEM (TPS); CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPICC CATHETER WITH SHERLOCK 3CG TIP POSITIONING SYSTEM (TPS); CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the purple lumen found leaking.Interestingly we did find that baxter needleless connector (10) (b)(4) leaked but when changed to (10) (b)(4) line was not leaking.It was checked the next morning ((b)(6)) and was ok then it was checked again on (b)(6) it did not leak with flush but nurse left and came back after about and hour and found a drop of fluid at connection of needleless connector and lumen.Additional information received 03/24/2022: what was infusion when the leak was discovered? cpn.Any patient symptoms related to fluid leak? no.Was the catheter removed? yes was a new catheter placed? exchanged (b)(6).What type of catheter securement was utilized? statlock.Was there any patient harm reported? no.
 
Additional Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of refx1855 showed two other similar product complaint(s) from this lot number.The complaints for this lot number (refx1855) have been reported from the same facility.
 
Additional Manufacturer Narrative
The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a leaking catheter was confirmed and the cause was manufacturing-related.The product returned for evaluation was one 5fr d/l powerpicc catheter.Light usage residues were observed throughout the sample and needleless injection caps were attached to the luer adapters.An attempt to infuse water through the sample using a 12ml syringe revealed the sample to be patent to infusion and aspiration; however, a leak was observed at the junction between the injection cap and the purple luer adapter.Inspection of the purple luer adapter using an iso taper gauge revealed that the gauge could not be fully inserted into the luer orifice.Microscopic inspection of the purple luer adapter revealed a protrusion in the luer material within the orifice.The mold defect appeared to prevent firm attachment of the cap, as well as full insertion of the taper gauge.A longitudinally aligned scoring mark was observed within the orifice adjacent to the mold defect.The molding defect observed within the purple luer orifice prevented firm attachment with the cap and resulted in leakage at the interface between the catheter and the cap.Photographs of the sample have been forwarded to the manufacturing site for further evaluation.
 
Event or Problem Description
It was reported the purple lumen found leaking.Interestingly we did find that baxter needleless connector (10) ur21i17075 leaked but when changed to (10) ur21k30026 line was not leaking.It was checked the next morning (3/22) and was ok then it was checked again on 3/23 it did not leak with flush but nurse left and came back after about and hour and found a drop of fluid at connection of needleless connector and lumen.Additional information received 03/24/2022: -what was infusion when the leak was discovered? cpn.-any patient symptoms related to fluid leak? no.-was the catheter removed? yes was a new catheter placed? exchanged 3/23.-what type of catheter securement was utilized? statlock.-was there any patient harm reported? no.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC CATHETER WITH SHERLOCK 3CG TIP POSITIONING SYSTEM (TPS)
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kelsey erickson
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key14054017
Report Number3006260740-2022-01212
Device Sequence Number4753196
Product Code LJS
UDI-Device Identifier00801741152573
UDI-Public(01)00801741152573
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K091324
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberCK000820
Device Lot NumberREFX1855
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/21/2022
Supplement Date Received by Manufacturer06/01/2022
Initial Report FDA Received Date04/08/2022
Supplement Report FDA Received Date06/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age75 YR
Patient SexMale
Patient Weight69 KG
-
-