Model Number H7493932800350
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event was estimated based on aware date as the event date was not provided.Pro code (product code): obj, itx.
 
Event or Problem Description
It was reported that device detachment occurred.The opticross oc35 peripheral imaging catheter was advanced for ultrasound examination of the target lesion.As the catheter was being used, the outer shaft separated.The case was able to be completed but the imaging was blurry.There was no harm to the patient.
 
Event or Problem Description
It was reported that device detachment occurred.The opticross oc35 peripheral imaging catheter was advanced for ultrasound examination of the target lesion.As the catheter was being used, the outer shaft separated.The case was able to be completed but the imaging was blurry.There was no harm to the patient.
 
Additional Manufacturer Narrative
B3 date of event was estimated based on aware date as the event date was not provided.Device evaluated by manufacturer: the product was returned for analysis.A kink was observed in the telescope assembly.The sheath was observed detached near the strain release section.During microscopic inspection noted pitting and degradation of the transducer housing consistent with saline damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTICROSS 35
Common Device Name
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14054359
Report Number2134265-2022-04091
Device Sequence Number14019502
Product Code OBJ
UDI-Device Identifier08714729984542
UDI-Public08714729984542
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K200733
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/20/2022
Device Model NumberH7493932800350
Device Catalogue NumberH7493932800350
Device Lot Number0028962616
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/21/2022
Supplement Date Received by Manufacturer06/01/2022
Initial Report FDA Received Date04/08/2022
Supplement Report FDA Received Date06/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1