| Model Number |
MCENDO550R |
| Medical Device Problem Code |
Optical Problem (3001)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
|
| |
|
Event or Problem Description
|
|
The user facility reported to terumo cardiovascular that during vein harvesting procedure, they noticed a foggy scope.It is unknown if the product malfunction caused or contributed to an injury.Terumo continues an attempt to gain more information regarding this event from the user facility.Product was changed out.Procedure was completed successfully with 2 minutes delay.
|
| |
|
Event or Problem Description
|
|
New information received.The product did not malfunction cause or contribute to an injury, even if temporary, or death.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on april 8, 2022.Upon further investigation of the reported event, the following information is new and/or changed: h3 (device evaluation anticipated by manufacturer - a second follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11.) h6: health effect - impact code: 2199 - no health consequences or impact.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: h6 (identification of evaluation codes 10, 3331, 3224, 19).Type of investigation #1: 10 - testing of actual/suspected device.Type of investigation #2: 3331 - analysis of production records.Investigation findings: 3224 - optical problem identified.Investigation conclusions: 19 - cause traced to user.Upon evaluation of the device the complaint was confirmed.The complaint sample was returned and visually inspected.A monocular was used to inspect the device and it was discovered that there was a crack on one of the internal lenses which resulted in a blurry visual field.A review of the device history record revealed no anomalies.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|
| |
|