| Model Number |
PDP423H |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient underwent a c-section procedure (b)(6) 2022.And suture was used.During the procedure, when the wound was closed, the surgeon had an unusual feeling at the first stich during suturing fat.The surgeon realized that the needle was broken at the swage part.An x-ray was performed to check for any broken needle remaining in the body, and the needle which was remaining in the fat layer was retrieved.The operation time was extended over 30 minutes.There were no patient consequences reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Component code: (b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained via: was there any change in the patient¿s post-operative care due to the prolonged unknown.After remove the needle, the patient is no problem.What is the lot number ? unknown.Trade name - irgacare®.Active ingredient(s) ¿ triclosan.Dosage form ¿ suture/solid/parenteral.Strength = 2360 g/m.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) date sent to the fda: 5/18/2022 this is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: d9 additional h3 investigation summary: a failed suture needle was submitted for fractographic evaluation.The component was identified as product code pdp423h.A fracture was observed at the suture attachment of the needle.The needle was received in two pieces, only one of the mating fracture surfaces was examined for this evaluation.A microscope was used to examine the fracture surfaces and surrounding area of the needle.The fracture surfaces were examined in multiple locations to determine the fracture mode.The evaluation revealed the fracture was composed of microvoid coalescence, which is evidence of a ductile fracture mode.Mechanical damage observed coincidental to the fracture provides additional evidence that the failure was induced by mechanical deformation leading to ductile overload.This was a ductile fracture.The evidence of this examination indicates that the breakage occurred at the attachment area of the needle during use due to tensile overload.There is no evidence of any material flaw or defect that would cause premature failure.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.The manufacturing records could not be reviewed as the batch number is unknown.¿ trade name - irgacare® ¿ active ingredient(s) ¿ triclosan ¿ dosage form ¿ suture/solid/parenteral ¿ strength ¿ = 2360 ¿g/m.
|
| |
|
Search Alerts/Recalls
|
|
|