• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR TISSUE EXPANDER; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR TISSUE EXPANDER; EXPANDER, SKIN, INFLATABLE Back to Search Results
Model Number 3504310M
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 03/16/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a female patient underwent a breast surgery with a 550cc mentor tissue expander and experienced a deflation on an unknown side post-operatively.As a result, the patient underwent explantation on (b)(6) 2022.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On april 22, 2022, mentor received additional information regarding the device serial number.The device serial number was updated to (b)(6).Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On (b)(6) 2022, mentor received additional information regarding the device date of implantation and device date of explantation.The device date of implantation has been updated to (b)(6) 2022.The device date of explantation has been amended to (b)(6) 2022.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On july 04, 2022, it was determined that the mentor failure analysis lab received the device for evaluation for this complaint on may 16, 2022.On july 08, 2022, the evaluation for the device was completed.Additional information states that the side of the implant was the right side.The patient's initials have been updated to s.N.In section a1 patient identifier.Device evaluation summary: visual analysis of the returned sample revealed that the exp rect remote 550cc breast implant was found to have a tear on the union between shell and patch.A microscopic examination was performed and the cause of the deflation could not be identified.Therefore a thickness measurement was conducted on the shell at the tear, and the thickness was within manufacturing specifications.Based on the information currently available, a product issue was identified during the investigation of the sample received.This product issue was addressed through mentor¿s quality system.During manufacturing, 100% of the devices were inspected for holes or tears.During the inspection, each device is inflated with air and submerged in 100% isopropyl alcohol (ipa).While submerged, a qualified operator will apply pressure on the device by gently squeezing and will observe for bubbles escaping the device that would indicate a leak.For the lot number reported in this complaint, no device was identified and rejected for leaks.In conclusion, with consideration to process controls and manufacturing records reviewed, the tear reported could not be confirmed as a manufacturing defect.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On july 14, 2022, mentor received additional information regarding the patient's date of birth and age.The patient's date of birth has been updated to (b)(6) 2016.The age of the patient has been updated to 6 years old.On july 27, 2022, mentor became aware that the the device evaluation summary in section h10 from the previous report erroneously referenced a breast implant when it should have been a tissue expander.The following device evaluation summary sentence has been updated from "visual analysis of the returned sample revealed that the exp rect remote 550cc breast implant was found to have a tear on the union between shell and patch." to "visual analysis of the returned sample revealed that the exp rect remote 550cc tissue expander was found to have a tear on the union between shell and patch." manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR TISSUE EXPANDER
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
michelleann garcia
3041 skyway circle north
irving, TX 75038
6465917981
MDR Report Key14059470
Report Number1645337-2022-03966
Device Sequence Number19988779
Product Code LCJ
UDI-Device Identifier00081317002468
UDI-Public00081317002468
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K842883
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/08/2023
Device Model Number3504310M
Device Catalogue Number3504310M
Device Lot Number7711996
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/17/2022
Supplement Date Received by Manufacturer04/22/2022
05/02/2022
07/04/2022
07/14/2022
Initial Report FDA Received Date04/08/2022
Supplement Report FDA Received Date05/11/2022
05/26/2022
07/08/2022
07/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/10/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age6 YR
Patient SexFemale
-
-