The customer reported that while preparing the subject device for an unspecified procedure, the knife wire broke but it was not energized and had not fallen off.Another device of the same model was used to complete the intended procedure.There was no patient or user injury reported due to the event.The device was returned to an olympus service center for evaluation.Upon inspection and testing of the returned device, it was observed that the broken part of the knife wire was burnt and it appeared that it had been broken by energization.This report is being submitted for the malfunction found during evaluation (the knife wire was burnt and broken by energization).Additional details have been requested regarding the reported event.At this time, no additional information has been provided.
|
|
Correction to g3 of the initial medwatch.The aware date should be 01-mar-2022.This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, possible causes may include: ·high frequency current was applied between the cutting wire and the tissue at the point of the contact.As a result, the current density at the contact area increased, and the cutting wire became instantly hot.·high frequency current was applied when the cutting wire and the tissue were being close to each other.As a result, an electrical discharge occurred, and the cutting wire became instantly hot.A definitive root cause cannot be identified.This issue is covered in the instructions for use: since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the cutting wire is very short, the output is too high or activated while the cutting wire touches metal parts of the endoscope, or the cutting wire is tightened too strong.When the cutting wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the cutting wire will be pushed out toward the papilla or move sideways.If the cutting wire breaks off, stop the output immediately and pull the slider completely to retract the broken cutting wire into the tube.Then withdraw the sphincterotome from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct and/or damage of the endoscope could result.Olympus will continue to monitor the field performance of this device.
|