• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC DAVINCI XI; PATIENT SIDE CART, 4-ARM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC DAVINCI XI; PATIENT SIDE CART, 4-ARM Back to Search Results
Model Number 380652-50
Device Problem Unintended System Motion (1430)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/09/2022
Event Type  malfunction  
Manufacturer Narrative
An intuitive surgical, inc.(isi) field service engineer (fse) contacted the customer to follow-up on the reported issue.The customer noted that the issue did not persist after contacting technical support.The fse checked the system logs and noted that there were no further issues in the logs.The customer also confirmed that there have been no other instances of the issue since the report of the event.A review of the site's complaint history does not reveal any related complaints involving this product or this event.No image or procedure video was provided for review.A review of the site's system logs for the reported procedure date was conducted by the tse.Investigation revealed the following possible related system errors: 22020|installed mcs (monopolar curved scissors/8x19) failed to engage on usm3 (wrist pitch axis failed to engage).This complaint is considered a reportable malfunction due to the following conclusion: during a procedure, the instrument exhibited unintended movement.Unintuitive motion could lead to subsequent tissue damage.While there was no harm or injury to the patient, the reported failure mode could likely cause or contribute to an adverse event if it were to recur.
 
Event Description
The robotics coordinator called an intuitive surgical inc.(isi) technical support engineer (tse) to report that during a case the previous day, the surgeon was working and the instrument moved backwards while they were using it.The robotics coordinator stated that this movement was unintended.The surgeon was able to complete the case, but wanted to have the tse review the logs.Tse noted error 22020 on universal surgical manipulator (usm3) during that timeframe.The robotics coordinator stated that it could have been a drape issue and they will monitor the system to see if the issue occurs again.Customer would like the field service engineer (fse) to follow up after next cases to ensure the issue did not return.The procedure was completed with no reported injury.Isi made multiple follow-up attempts to obtain additional information.However, no further details have been received as of the date of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAVINCI XI
Type of Device
PATIENT SIDE CART, 4-ARM
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key14067537
MDR Text Key298580586
Report Number2955842-2022-10969
Device Sequence Number1
Product Code NAY
UDI-Device Identifier00886874110720
UDI-Public(01)00886874110720
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131861
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 03/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number380652-50
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/10/2022
Initial Date FDA Received04/09/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/24/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES
-
-