BOSTON SCIENTIFIC CORPORATION INVIVE; PULSE GENERATOR, WITH CARDIAC RESYNCHRONIZATION (CRT-P) Back to Search Results
Model Number V173
Device Problems Premature Discharge of Battery (1057); Pacing Problem (1439); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problems Laceration(s) (1946); Syncope/Fainting (4411); Asystole (4442)
Event Date 02/05/2022
Event Type  Injury  
Event Description
It was reported that the patient with this cardiac resynchronization therapy pacemaker (crt-p) presented to the emergency room after experiencing syncope and falling, causing a laceration to their head.It was noted this was due to a 30 second period of asystole.The device was found to be operating in safety mode.The device exhibited brady pacing not delivered when required and premature battery depletion.A temporary pacing wire was placed, then the device was explanted and replaced.No additional adverse patient effects were reported.At this time, the product has not been returned.If the product is returned, analysis will be performed, and this report would be updated at that time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVIVE
Type of Device
PULSE GENERATOR, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key14067539
MDR Text Key289078768
Report Number2124215-2022-05336
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526536632
UDI-Public00802526536632
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S079
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/09/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/24/2014
Device Model NumberV173
Device Catalogue NumberV173
Device Lot Number101940
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/05/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/27/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age73 YR
Patient SexFemale
-
-